看一级毛片**直播在线_综合在线免费视频_国语对白嫩模_国产精品区一区二区三,午夜精品视频在线观看美女,夜恋秀场全免费视频人成,亚洲成a人无码亚洲成a无码网址

設為首頁 加入收藏 聯系我們

公司動態
當前位置:首頁 > 公司動態 > icp光譜儀怎么操作(ICP spectrometer how to operate)
icp光譜儀怎么操作(ICP spectrometer how to operate)
點擊次數:0 發布時間:2016-10-04
 

icp光譜儀是目前常用的分析儀器,它應用在各個領域,用戶在使用過程中,多數不了解icp光譜儀怎么操作,icp光譜儀正確操作方法是什么?下面讓我們來了解一下:

ICP spectrometer is a commonly used analytical instrument, it is used in various fields, users in the process of use, most do not know how to operate the ICP spectrometer, ICP spectrometer correct operation method is what? Let's get to know:

1 開機順序:實驗室總電源→氬氣→冷卻水→抽風機→儀器主機→進樣泵→電腦→smart analyzer vision工作站

1 boot order: total power, argon, laboratory, exhaust fan, cooling water, injection pump, instrument host computer, smart analyzer vision workstation

2 方法編輯:點擊smart analyzer vision工作站左邊工具箱中的method按鈕進入方法編輯窗口,點擊new method,輸入方法名,ok↓確定。

2 methods of editing: click on the smart vision workstation analyzer left in the toolbox method button to enter the method of editing window, click on the new method, the input method name, OK: sure.

進行測量

To measure

當方法編輯好后,就可以進行點火開始測量。點火:點擊system/setup devices,分別設置等離子氣體流量為10l/min、輔助氣流量為1l/min、霧化器氣流量為1l/min,點擊apply 進行沖洗氣路2分鐘。在矩管圖中觀察以上各氣流量回零后,點擊plasma點燃等離子體,點火后點擊pump啟動進樣泵。在進樣前吸入蒸餾水或空基體溶液5∽15分鐘,以加熱和平衡組成成分。

When the method is edited, the ignition can be started. Ignition: click on the system/setup devices, were set up plasma gas flow rate of 10l/min, auxiliary gas flow 1l/min, nebulizer gas flow rate of 1l/min, click apply to wash gas for 2 minutes. In the torque tube chart above the gas flow back to zero, click plasma to ignite the plasma, ignition after the click pump to start the sample pump. In the sample before inhalation of distilled water or 5 ~ 15 minutes free matrix solution, composition for heating and balance.

4、數據處理

4, data processing

5、關機,順序關機:關氬氣→冷卻循環水→儀器主機→抽風機→電腦→實驗室總電源。

5, shutdown, shutdown order: argon, cooling water, exhaust fan, computer, instrument, laboratory of general power.

以上就是小編為您具體介紹的icp光譜儀怎么操作,對于以上內容您是否了解了呢,正確操作icp光譜儀如上所述,希望對您有所幫助。

The above is a small series for your specific ICP spectrometer how to operate, for the above content you understand it, the correct operation of the ICP spectrometer as mentioned above, I hope to help you.


聯系人:李先生
電話:010-67892001
手機:13521683201
點擊這里給我發消息
點擊這里給我發消息
主站蜘蛛池模板: 辽源市| 天峨县| 花莲市| 西乌珠穆沁旗| 崇州市| 沙坪坝区| 台中县| 井研县| 海城市| 连州市| 延边| 大田县| 乃东县| 华池县| 鄂托克前旗| 平谷区| 兴和县| 金寨县| 霞浦县| 平塘县| 富蕴县| 霸州市| 河北区| 类乌齐县| 塔城市| 黔东| 镇安县| 江永县| 定襄县| 东城区| 宜宾市| 南陵县| 二连浩特市| 富阳市| 桑日县| 都昌县| 天水市| 广河县| 东兰县| 通道| 渑池县|